Bit-Exact Reconstruction and Certification of Electrocardiogram Waveforms from Vector-Encoded PDF Files

Reconstruct and certify bit-exact waveforms from GE MUSE 12-lead ECG PDFs.

badge Open in MATLAB Online
badge badge Binder
badge
badge

Jonathan W. Waks, MD
Harvard-Thorndike Electrophysiology Institute, Department of Cardiovascular Medicine,
Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, USA
braveheart.ecg@gmail.com


image image

load_musepdf parses a GE MUSE 12-lead ECG PDF generated with vector graphics, extracts the polylines that make up the waveforms, and inverts the double quantization introduced when the device rendered the ECG for print. The result is a bit-exact reconstruction of the original waveforms which were used to generate the PDF rather than an approximation. This allows use of PDF ECGs in AI-ECG or vectorcardiographic analyses without needing to worry that small errors in signal digitization could cause unpredictable downstream errors when compared to use of raw data formats such as XML or DICOM.

As the parameters that are used for reconstruction (such as the stream unit to analogue-to-digital [ADU] ratio or baseline zero voltage references) are not directly encoded in the PDF (unlike raw digital formats such as XML or DICOM), and must be extracted or presumed, the software leverages the double quantization of PDF stream unit values to provide certification that bit-exact reconstruction was successful. This is critically important to avoid propagating silent errors into downstream analyses when processing PDFs from unknown sources.

The software also identifies signal clipping, where part of the waveform is not rendered on the PDF due to being outside of the rendering area. This results in loss of data which cannot be recovered. The ECGs flagged for clipping should be inspected prior to use as they can contain significant reconstruction errors.

Please see our preprint at medRxiv for a full discussion of the software methods and results of testing on a large database of PDF ECGs in multiple formats.


Supported formats

Property Supported
Source GE MUSE-generated PDFs only
Sweep speed 25 mm/s
Gain 5, 10 and 20 mm/mV, uniform across all leads
Layouts 12×1 rhythm; 4×3 2.5 s with 0, 1 or 3 rhythm strips; 6×2 5 s
Pages 1, or 2 where the second page carries no waveforms
Compression FlateDecode with a direct /Length

Not supported: 50 mm/s and 12.5 mm/s sweep speeds, mixed limb/precordial gains, non-MUSE PDFs, and PDFs where the waveform streams use an indirect /Length reference. Each of these is detected and raises a descriptive error rather than returning wrong data.

Only the 12×1 layout aligns all leads in time. In 4×3 and 6×2 formats the leads are recorded in different time windows, which matters if you intend to build median beats.


Requirements

Two implementations are provided and produce identical output:

Implementation Requirements
matlab/load_musepdf.m MATLAB R2022a or later
python/load_musepdf.py Tested with Python 3.14.3 with numpy 2.4.5

Usage

Both implementations require input of a PDF filename as a string followed by various options.

Sample PDF and matching XML files are provided to test the software. XML files can be read using load_musexml.m and load_musexml.py.

pdf2ecgdemo.m and pdf2ecgdemo.py can be run to load an ECG in PDF and XML formats and then compare the outputs.

Options

Option Effect
verbose Print parsing, layout, calibration, and certification detail
adu Return signals in integer ADUs instead of mV
exportfile Write the processed ECG data to <basename>_digitized.csv (or _digitized_adu.csv)
calcleads (default) Extracts leads I, II and V1–V6; computes leads III, aVR, aVL and aVF from leads I and II
pdfleads Extract all leads directly from the PDF

MATLAB

Open in MATLAB Online

% Process PDF files with different options
[signals, info] = load_musepdf('ecg.pdf');
[signals, info] = load_musepdf('ecg.pdf', 'verbose');
[signals, info] = load_musepdf('ecg.pdf', 'adu', 'pdfleads', 'exportfile');

% Report total number of clipped samples
% If 0 then there is no clipping
signals.totalClipping

% Report if ECG passed certification tests
% If 1 then passed
signals.certified
% PDF data
[signals, info] = load_musepdf('ecg.pdf','verbose');
II_pdf = signals.II;

% XML data
[hz, I, II, V1, V2, V3, V4, V5, V6] = load_musexml('ecg.xml');
II_xml = II';    % Convert to column vector

% Show that XML and PDF leads are exactly equal
% Be alert for floating point errors as we are comparing mV
exact = isequal(II_pdf,II_xml);
fprintf('PDF and XML leads are equal: %s\n', string(exact));

% Plot XML and PDF leads
figure
hold on
title('Comparison of XML and Reconstructed PDF Waveforms')
plot(II_xml, 'k', 'LineWidth', 2 ,'DisplayName','XML')
plot(II_pdf, 'r--', 'LineWidth', 1 ,'DisplayName','PDF')
xlabel('Samples')
ylabel('mV')
legend

Python

Binder

# Process PDF files with different options
from load_musepdf import load_musepdf

signals, info = load_musepdf("ecg.pdf")
signals, info = load_musepdf("ecg.pdf", "verbose")
signals, info = load_musepdf("ecg.pdf", "adu", "pdfleads", "exportfile")

# Report total number of clipped samples
# If 0 then there is no clipping
print(signals["totalClipping"])

# Report if ECG passed certification tests
# If 1 then passed
print(signals["certified"])
from load_musepdf import load_musepdf
from load_musexml import load_musexml
import numpy as np
import matplotlib.pyplot as plt

# PDF data
signals, info = load_musepdf("ecg.pdf",'verbose')
II_pdf = signals["II"]
hz_pdf = signals["hz"]

# XML data
hz_xml, I_xml, II_xml, V1_xml, V2_xml, V3_xml, V4_xml, V5_xml, V6_xml = load_musexml("ecg.xml")

# Show that XML and PDF leads are exactly equal
# This will fail if the leads are of different lengths too
exact = np.array_equal(II_pdf, II_xml)
print("PDF and XML leads are equal:", exact)

# Plot XML and PDF leads
plt.figure(figsize=(12, 4))
plt.plot(II_xml, "k-", linewidth=2, label="XML")
plt.plot(II_pdf, "r--", linewidth=1, label="PDF")
plt.xlabel("Samples")
plt.ylabel("Amplitude (mV)")
plt.title("Comparison of XML and Reconstructed PDF Waveforms")
plt.legend()
plt.grid(True, alpha=0.3)
plt.show()

There is also a command line entry point:

python load_musepdf.py ecg.pdf verbose exportfile

Where MATLAB returns structs, Python returns dictionaries with the same field names. Lead arrays are 1-D numpy arrays.

calcleads vs pdfleads

MUSE computes leads III, aVR, aVL, and aVF from leads I and II internally, rounds them to integer ADU values, and these ADU values are then used to generate the PDF. pdfleads recovers exactly those values. calcleads instead calculates leads III, aVR, aVL, and aVF from leads I and II, resulting in values that can contain half-integer ADU values due to division by 2.

In calcleads mode the derived leads inherit the clipping and certification of leads I and II. If both source leads are bit-exact then anything derived from them is too, and if either lead has clipping, the derived leads also inherit that error.


Output

signals

Field Contents
I, II, III … V6 Lead data, mV by default or ADU with the adu option
hz Sampling frequency
units Amplitude units of the lead signals in 'mV' or 'ADU'
totalClipping Total clipped samples across all 12 leads
certified 1 pass, 0 fail, NaN indeterminate
Optional fields if rhythm strip is present
rhythm Rhythm strip leads, for layouts that contain them, associated with a field of its name (e.g. rhythm.II)
totalClippingRhythm Total clipped samples across all rhythm strip leads

info

Field Contents
hz Sampling rate in Hz.
leadOrder Lead names in the order used throughout the output.
leadsMode Mode for augmented leads ('calcleads' or 'pdfleads').
ecgFormat Page layout of the source ECG (e.g. '12×1').
ecgFormatStr Additional description of the PDF layout.
numPolylines Total number of vector polylines extracted.
numLeadsDetected Number of ECG leads detected.
cmMatrix Affine transform mapping source coordinates to centimeters.
certified Flag for if passed (1) or failed (0) certification.
diagnostics Structure containing results of validation tests. (see below)
lowPassFreq Low-pass filter cut-off frequency in Hz.
XMLmicrovoltsLSB μV per ADU (value used is assumed from MUSE).
unitsPerMv Stream units per millivolt.
mmPerMv ECG amplitude gain (mm per mV).
mmPerSec ECG sweep speed (mm per second).
alpha Stream-to-ADU conversion, α.
offsets Per-lead baseline offset for each lead in stream units.
numPages Number of pages in the source PDF.
numStreams Number of data streams.
clipping Nested structure with per-lead clipping details. For each lead includes total number of clipped samples and which samples are clipped. Includes rhythm strips if present.
maxStreamUnits Per-lead maximum value in stream units.
minStreamUnits Per-lead minimum value in stream units.
Optional fields if rhythm strip is present
rhythmStripLeads Rhythm strip lead names.

Diagnostics

The info.diagnostics structure contains information on certification test results for each lead.
The test results are explained in detail in the methods manuscript.

Field Contents
certified Overall certification result: passed (1), failed (0), or indeterminate (NaN).
W Result of the residue width test. Note: calculated leads return NaN
WMargin How much of the expected residue arc is filled. 0 is completely filled. Positive values indicate the arc is not completely filled, and negative values indicate values outside the arc (test failed).
passWLimit The maximum value of W to pass the residue width test. W > passWLimit fails.
passW Result of the residue width (W) test: passed (1), failed (0), or indeterminate (NaN)
GCD Greatest common divisor of the stream unit values. Values > 1 fail.
numD Number of distinct differences used in the GCD test.
passGCD Result of the GCD test: passed (1), failed (0), or indeterminate (NaN)
BOut Number of residues falling outside the expected window. Values > 0 fail
BOutClasses Number of distinct residue classes among those outliers in BOut
passB Result of the baseline test: passed (1), failed (0), or indeterminate (NaN).
Optional fields if in calcleads mode
calcleadsOriginal The above values prior to any adjustment of leads III, aVR, aVL, and aVF due to calcleads mode. In calcleads mode leads III, aVR, aVL, and aVF inherit certification from leads I and II.
Optional fields if rhythm strip is present
rhythm Same data as above for rhythm strips.

Important Information

  • Always check certified before using the output quantitatively. A value of 0 means at least one lead is provably not bit-exact, most often because of clipping.
  • Clipping is flagged from a single sample crossing a set rendering threshold. There is no way to tell whether such a sample merely touched the limit or was cut off far above it, and the error due to clipping therefore cannot be determined from the PDF alone.
  • 4.88 µV/LSB is assumed, matching the MUSE XML. Due to the certification tests, a device configured differently would fail certification rather than silently return incorrect data.

Citation

If you use this software, please cite:

Abdelrahman, K., Stabenau, H. F., Patlatzoglou, K., Sau, A., Zeidaabadi, B., Pastika, L., Peters, N. S., Ng, F. S., Kramer, D. B., & Waks, J. W. Bit-exact reconstruction and certification of electrocardiogram waveforms from vector-encoded PDF files. medRxiv. 9/2026. https://doi.org/10.64898/2026.09.23.26362281

BibTeX:

@article{abdelrahman2026pdf2ecg,
  title   = {Bit-Exact Reconstruction and Certification of Electrocardiogram Waveforms from Vector-Encoded PDF Files},
  author  = {Abdelrahman, K. and Stabenau, H. F. and Patlatzoglou, K. and Sau, A. and Zeidaabadi, B. and Pastika, L. and Peters, N. S. and Ng, F. S. and Kramer, D. B. and Waks, J. W.},
  journal = {medRxiv},
  year    = {2026},
  doi     = {10.64898/2026.09.23.26362281},
  note    = {Preprint}
}

License

Copyright 2026 Jonathan W. Waks All rights reserved. This program is free software: you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation, either version 3 of the License, or (at your option) any later version. This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General Public License for more details. You should have received a copy of the GNU General Public License along with this program. If not, see https://www.gnu.org/licenses/ or the LICENSE file included in this repository.

About

Bit-Exact Reconstruction and Certification of Electrocardiogram Waveforms from Vector-Encoded PDF Files