Get glucose data from the Medtronic MiniMed Go / Instinct sensor into xDrip+, Nightscout, and smartwatches.

If you have an Instinct sensor (made by Abbott, sold by Medtronic) and xDrip+'s CareLink Follower returns HTTP 204 with no data — this repo explains why and fixes it.


The problem

The Instinct sensor and the MiniMed Go app do not use the legacy CareLink API that existing tools target. They use a newer API that Medtronic's discovery config calls baseUrlCumulus.

xDrip+ has not implemented it. From the xDrip discussion:

"The instinct sensor seem to use a new api called Cumulus so these sensors probably won't work until this is implemented."

As of August 2026 that is still the case. Every existing bridge (carelink-bridge, minimed-connect-to-nightscout) targets the old endpoints, so Instinct users get nothing.

Additionally, the MiniMed Go app posts no persistent notification containing the glucose number (its alarms only say "high"/"low"), so xDrip+'s Companion App mode has nothing to scrape either. Juggluco cannot take over the sensor because it was not activated with Abbott's own Libre 3 app — and running two apps against a Libre 3 in parallel breaks both connections, which would compromise the official clinical alarms.

That leaves the cloud API as the only viable path.

The finding

POST {baseUrlCumulus}/display/message enforces a minimum client version. Send an old one and the server replies:

HTTP 206
{"message": "Upgrade application"}

xDrip+ was sending appVersion: "3.6.0", which Medtronic no longer accepts.

Tested against a live MiniMed Go + Instinct account (EU, care partner), varying one field at a time:

appVersion result
3.5.0, 3.6, 3.6.1, 3.6.9 206 Upgrade application
3.7, 3.7.0, 3.8, 3.8.0, 4.0.0, 10.0.0 200 OK

Boundary: ≥ 3.7 accepted, ≤ 3.6.9 rejected. Both X.Y and X.Y.Z forms work.

The working request:

POST https://clcloud.minimed.eu/connect/carepartner/v13/display/message
Authorization: Bearer <access_token>
Content-Type: application/json

{
  "username":   "<care partner username>",
  "role":       "carepartner",
  "patientId":  "<patient username>",
  "appVersion": "3.8.0"     // <-- must be >= 3.7
}

appVersion is a body field, not a header. Twelve header variants were tested and all failed: User-Agent spoofing (CareLinkConnect, okhttp, com.medtronic.carepartner) and the headers x-app-version, app-version, client-version, x-client-version and x-carelink-app-version.

os: "android" is optional. An earlier version of this document claimed it was required — it is not. A control request with appVersion: "3.8.0" and no os field returns 200 with full data. The real app does send it, so this bridge includes it, but it is not what unblocks the request.

The two version fields behave differently

Where Constraint
Discovery path /discover/android/<ver> 3.5 → Cumulus v11; 3.63.8v13; 3.9+ falls back to a legacy config (/connect/carepartner/v2, no Auth0)
Body appVersion must be ≥ 3.7, no upper bound

So the discovery URL must stay at 3.8 while the body value can go higher. If Medtronic raises the minimum again, only the body needs changing.

Response format differences

  1. Everything is nested under patientData (the old API had it at the root).
  2. Readings use timestamp, not datetime, and carry no timezone — they are the patient's local time (patientData.clientTimeZoneName).
  3. The sgs array is not sorted by time.

Architecture

Instinct sensor (Abbott)
   │ BLE
   ▼
MiniMed Go app  ──uploads──▶  CareLink cloud
                                   │  this bridge polls /display/message
                                   ▼
                        Nightscout-compatible API
                                   │
                         xDrip+ (Nightscout Follower)
                                   │
                    WatchDrip+ ──▶ smartwatch

The bridge is read-only against the cloud. It never touches the sensor or the official app, so the official clinical alarms keep working untouched.

Measured latency

Hop Delay
Sensor → MiniMed Go → CareLink cloud ~4.5 min (not controllable)
Cloud → bridge 0–POLL_INTERVAL
Bridge → xDrip+ xDrip's follower interval
xDrip+ → watch seconds

Realistic end to end: 5–8 minutes. The sensor itself produces a reading every minute.


Setup

Requires a CareLink care partner account (a patient account returns HTTP 204), approved by the patient in MiniMed Go → Profile → CareLink → Manage care partners.

npm install
npm run login    # opens a browser at Medtronic's real login page
npm run probe    # verifies the API returns your data
npm run serve    # starts the Nightscout-compatible server

npm run login writes logindata.json (access + refresh token). Treat it as a credential — it grants access to the patient's health data. It is gitignored.

Environment variables

Variable Required Description
LOGINDATA_JSON in the cloud Contents of logindata.json, one line. Seeds the token.
API_SECRET recommended Protects reads. Without it the endpoint is public.
MMCONNECT_SERVER no EU (default) or US
POLL_INTERVAL no Seconds between CareLink polls. Default 120. Do not go below 60.
CARELINK_PATIENT no Only if the care partner follows several patients
DATA_DIR no Persistent dir for the refreshed token. Use a volume.
PORT no Injected by most hosts

Persist DATA_DIR on a volume. Refresh tokens rotate; without persistence a redeploy falls back to the seed token, which may already be invalid.

⚠️ Only one instance may refresh the token

Auth0 rotates refresh tokens, and reusing a spent one triggers reuse detection, which revokes the whole token family.

In practice: if the bridge is deployed and you also run something locally that refreshes using an older copy of logindata.json, you kill the deployment. That happened while building this — 13 consecutive 403s and no data on the watch until a fresh login.

Two safeguards are built in:

  • The diagnostic probes never refresh. If the local access token is expired they fail with instructions instead of touching anything.
  • The server self-heals: if its stored refresh token is rejected and LOGINDATA_JSON now holds a different one, it adopts the seed and persists it. So after a fresh login you only need to update that variable — no need to clear the volume by hand.

Endpoints

Route Purpose
GET /health Diagnostics: cache age, reading count, last value, errors. No auth.
GET /api/v1/status.json Nightscout status
GET /api/v1/entries.json?count=N Readings, newest first. Alias /api/v1/entries/sgv.json

Auth: api-secret header (SHA1 of the secret, Nightscout convention) or ?token=.

Deploying

See DEPLOY.md (Railway) and SELFHOST.md (Docker / systemd / Tailscale). A Dockerfile and docker-compose.yml are included.

The container runs as root on purpose: hosts commonly mount volumes owned by root, and dropping to an unprivileged user makes the token write fail with EACCES.

Client setup

xDrip+: Hardware Data Source → Nightscout Follower, URL https://<API_SECRET>@your-host, then Inter-app settings → Broadcast Service API = ON.

WatchDrip+ (Xiaomi Smart Band 8 Pro / 9 / 10, Redmi Watch 4): Enable service ON, Enable Xiaomi service ON, Enable device OFF. Watchfaces "Watchdrip+ Graph" and "Watchdrip+ Big" live in the modded Mi Fitness app under Mods → english category.


Credits

The appVersion/os requirement documented here was found by experimentation and, as far as we can tell, was not published anywhere before. PRs welcome — especially to get this into xDrip+ so the bridge becomes unnecessary.

Disclaimer

Not a medical device. Not certified. Secondary visualisation only.

Keep your clinical alarms in the official app. Never dose insulin based on what this shows. Data arrives minutes late by design. This is a community project with no warranty of any kind — see LICENSE.


En español

Este proyecto resuelve el caso de los sensores Instinct de Medtronic (fabricados por Abbott), que no funcionan con el CareLink Follower de xDrip+ porque usan una API distinta llamada Cumulus.

La clave: POST /display/message rechaza versiones de cliente antiguas. xDrip+ mandaba appVersion: "3.6.0" y el servidor exige ≥ 3.7; si no, responde 206 {"message":"Upgrade application"}. Va en el cuerpo, no en cabeceras (probadas doce variantes). El campo os es opcional.

El bridge lee solo de la nube: no toca el sensor ni la app oficial, así que las alarmas clínicas siguen intactas. Latencia real medida: 5–8 minutos.

Necesitas una cuenta de care partner (con la de paciente da HTTP 204), aprobada desde MiniMed Go → Perfil → CareLink → Gestionar cuidadores.

Visualización secundaria, no certificada. Nunca dosifiques insulina con esto.

About

Get glucose from Medtronic MiniMed Go / Instinct sensors into xDrip+, Nightscout and smartwatches. Implements the CareLink Cumulus API that xDrip+ lacks (HTTP 204 fix).

Topics

Resources

Stars

2 stars

Watchers

0 watching

Forks

Releases

Packages

Contributors

Languages