kardia-rs

kardia-rs is a Rust reverse-engineering workbench for capturing, inspecting, and decoding the Bluetooth Low Energy ECG stream from an AliveCor KardiaMobile 6L.

The verified path connects to a bonded Kardia 6L, records every raw notification, displays a live six-limb-lead view, inspects the observed packet transport, exports confirmed M2 recordings to analysis-friendly CSV, and renders printable six-lead ECG reports as vector PDF or SVG.

Warning

This is independent research software, not a medical device. It is not affiliated with or endorsed by AliveCor, and its output must not be used for diagnosis or treatment. Report amplitude uses a specification-derived, provisional scale; signal polarity and physical-unit calibration have not been independently validated. Automated intervals and axes are experimental and have not been validated against an annotated clinical ECG database. The optional ML rhythm-similarity model is trained on public non-Kardia data and has not been clinically or device-domain validated.

Contents

Status

Tested with a KardiaMobile 6L running firmware 3.0.1 on macOS 15.7.4:

Capability Status
BLE scan, connection, GATT discovery, and characteristic reads Confirmed
Encrypted/bonded vendor characteristic access Confirmed
Compatibility command generation Confirmed on a normally bonded device
M2 dual-channel stream 36-byte packets, 9 interleaved i16 pairs, 300 Hz/channel
Channel identity Channel 1 = lead I; channel 2 = lead II
Live I/II/III/aVR/aVL/aVF display Working in uncalibrated raw counts
Raw capture inspection and M2 CSV export Working
A4 six-lead ECG report Vector PDF and SVG at standard 25 mm/s and 10 mm/mV
Automated measurements Experimental HR, PR, QRS, QT/QTcB/QTcF, and device-native P/QRS/T axes
Optional ML report panel Three-way rhythm similarity with conservative confidence/margin abstention
M4 dual-lead 600 Hz request Command sent, but observed transport remains M2-compatible at 300 Hz
Signal polarity Device-native direction remains unverified; report supports manual inversion
Volts-per-count Provisional inference: 10 mV / 65536 per stored count; simulator validation pending
Linux and Windows BLE capture Architecturally supported, not yet device-tested

The sanitized protocol evidence log is in docs/re/kardia-6l.md. Public evidence and redaction boundaries are documented in docs/re/protocol-provenance.md.

Features

  • Discovers Kardia-like BLE advertisements and inspects GATT services.
  • Builds and sends the device compatibility command only after the operating system establishes a bonded, encrypted link.
  • Journals connection, bonding, subscription, unlock, and collection stages.
  • Preserves raw command indications and ECG notifications before decoding.
  • Displays a throttled six-lead view while lossless M2 capture continues.
  • Distinguishes requested/nominal mode metadata from observed packet behavior.
  • Inspects packet sizes, cadence, command indications, and M2 compatibility.
  • Exports M2 samples with both raw channels and all six derived limb leads.
  • Renders simultaneous I/II/III/aVR/aVL/aVF strips on a physically dimensioned A4 ECG grid as PDF or SVG.
  • Builds an aligned median complex and prints confidence-gated experimental rate, interval, QT-correction, and frontal-axis measurements.
  • Optionally runs a local, versioned ONNX waveform model and prints sinus-rhythm-like, AF-like, or other/noisy similarity only when conservative quality, confidence, and class-separation gates pass.
  • Keeps captures out of Git by default because they may contain health data and device identifiers.

Requirements

  • Rust 1.82 or newer
  • A KardiaMobile 6L
  • Bluetooth Low Energy support
  • An encrypted/bonded host-device relationship

Only macOS/CoreBluetooth has completed the full live-capture path so far.

Installation

git clone https://github.com/ShaneBreazeale/kardia-rs.git
cd kardia-rs
cargo build --release

During development, replace target/release/kardia with cargo run -p kardia-cli -- in the examples below.

Quickstart

Close the AliveCor mobile app before connecting so it does not claim the device. Wake the Kardia, then scan:

cargo run -p kardia-cli -- scan --seconds 10

Capture 30 seconds in confirmed M2 mode and show live labeled output:

cargo run -p kardia-cli -- capture-raw \
  --mode m2 \
  --live \
  --rescan 20 \
  --seconds 30 \
  --out captures/kardia-m2.csv

The capture command persists every raw packet first, updates the live view at a lower terminal rate, and prints an observed-transport inspection when it finishes.

Inspect any saved capture again:

cargo run -p kardia-cli -- inspect-raw captures/kardia-m2.csv

Export a confirmed M2 capture:

cargo run -p kardia-cli -- export-six-lead-m2 \
  captures/kardia-m2.csv \
  --out captures/kardia-six-lead.csv

The exported columns include the raw channel pair, leads I and II, derived III/aVR/aVL/aVF, an exact 300 Hz sample clock, and the original packet receive timestamp. Values remain device-native counts, not millivolts.

Render the first 10 seconds as a printable report:

cargo run -p kardia-cli -- render-ecg \
  captures/kardia-m2.csv \
  --out output/pdf/kardia-six-lead.pdf

Six-lead ECG Reports

render-ecg accepts only confirmed M2 captures and draws six simultaneous limb leads: measured I and II plus derived III, aVR, aVL, and aVF. It does not invent V1-V6 and labels their absence on the page.

Sample Report

An anonymized six-lead limb ECG rendered on a standard ECG grid

This anonymized 9.96-second sample was rendered directly from a device capture. Open the full-size vector report. The waveform and automated measurements are research output, not a medical interpretation.

The defaults produce a landscape A4 page with:

  • 25 mm/s paper speed and 10 mm/mV gain;
  • a 1 mm minor and 5 mm major ECG grid;
  • a 1 mV by 200 ms calibration pulse for every strip;
  • an exact 300 Hz horizontal sample clock;
  • per-lead median subtraction for vertical placement; and
  • no smoothing, notch, high-pass, or low-pass filter applied to the displayed waveform.

The report also includes an experimental automated measurements panel:

  • ventricular rate from the first-to-last detected QRS span;
  • PR interval, global QRS duration, and global QT interval from an aligned representative median complex;
  • Bazett (QTcB) and Fridericia (QTcF) rate correction using average RR; and
  • frontal P/QRS/T axes calculated from the two independent I/II channels.

Beat detection uses a separate analysis copy with baseline suppression, smoothing, multi-channel slope energy, and a refractory period. Qualifying beats are aligned and combined sample-by-sample with a median. Wave boundaries are then estimated from simultaneous channel slopes and return-to-baseline tests. These analysis operations never alter the waveform drawn on the report.

Values that fail technical confidence checks are printed as --. “Technical quality” describes detector confidence only, not whether an ECG is medically normal. The P/QRS/T axes carry an asterisk because they retain the currently unverified device-native polarity. The measurements must not be used for diagnosis and still require validation against annotated reference databases.

The default report interval is 10 seconds. Select a later section of a longer capture with --start-seconds:

cargo run -p kardia-cli -- render-ecg \
  captures/kardia-m2.csv \
  --start-seconds 10 \
  --seconds 10 \
  --out output/pdf/kardia-six-lead-page-2.pdf

Use an .svg output extension for editable vector output. --speed-mm-s, --gain-mm-mv, and --mv-per-count are available for controlled experiments; --invert reverses device-native polarity.

The default voltage conversion is the explicit hypothesis 10 mV / 65536 = 0.000152587890625 mV per stored signed count. AliveCor specifies a 10 mV peak-to-peak input range and 14-bit resolution, while every observed value has its low two bits clear. This is consistent with a 14-bit sample left-aligned in a signed 16-bit field, but it is not yet an electrical calibration. Every report carries a visible provisional-amplitude warning until an isolated ECG simulator or equivalent traceable reference confirms the scale and polarity.

Optional ML Rhythm Similarity

render-ecg can optionally load a reviewed model manifest. Without --model, no ML runtime or model is used:

cargo run -p kardia-cli -- render-ecg \
  captures/kardia-m2.csv \
  --model models/limb6-rhythm-v0.2.0.json \
  --out output/pdf/kardia-six-lead-with-ml.pdf

The model receives a separate analysis copy of the same six displayed limb leads. The 300 Hz signal is reduced to the model's declared 100 Hz input, median-centered, normalized using the two independent I/II channels, and clipped to the manifest's training range. This processing does not alter the waveform drawn on the report.

Before inference, the CLI validates the model SHA-256, manifest schema, label order, lead order, tensor shape, research-only declaration, and threshold ranges. It abstains when deterministic technical quality is not GOOD, the top probability is below its class-specific threshold, or the top-two class margin is too small. The report always labels the result as research-only and the Kardia domain as unvalidated.

The automated summary keeps deterministic observations separate from the ML decision. It can still report rate, RR variability, P-before-QRS detection on the median beat, PR, QRS, and QTcF when the rhythm classifier abstains.

The three outputs are similarity categories, not diagnoses:

  • sinus-rhythm-like: resembles public training records explicitly labeled sinus rhythm;
  • af-like: resembles records explicitly labeled atrial fibrillation; and
  • other/noisy: alternative rhythms, listed signal contamination, or other unsupported patterns.

“Sinus-rhythm-like” does not mean “normal ECG,” and an abstention is an intentional output rather than an error. V1-V6-dependent findings must not be inferred from this device. The complete reproducible data-labeling, training, calibration, evaluation, and ONNX-export workflow is in ml/README.md. Architecture, thresholds, held-out results, and device-domain limitations are recorded in the v0.2 model card. The more restrictive v0.1 model card is retained for comparison.

Pairing and Troubleshooting

The ac06000x vendor characteristics require an encrypted BLE link. Without a valid bond, writes can return ATT error 5 (Insufficient Authentication) and vendor notification subscriptions can stall.

If capture does not reach the unlock write:

  1. Close the AliveCor app.
  2. Forget or remove the Kardia relationship from the phone.
  3. Remove the Kardia battery for about 15 seconds, then reinstall it.
  4. Wake the device and rerun capture-raw.
  5. Accept the macOS pairing dialog when it appears.

Run the lower-level probes when isolating a failure:

Inspect every service and characteristic:

cargo run -p kardia-cli -- gatt-dump --rescan 20

Read every readable characteristic:

cargo run -p kardia-cli -- read-probe --rescan 90 --verbose

Compare a standard battery subscription with the vendor ECG subscription:

cargo run -p kardia-cli -- notify-probe --rescan 90 --verbose

The device advertises for a short window. Wake it immediately before scanning, and prefer capture-raw first when collecting evidence.

Capture Format

Raw captures are append-only CSV-like text:

 # kardia_raw_capture_v2 ... requested_mode=M2 nominal_sample_rate_hz=300 ...
 # stage unix_micros=... name="connect_and_discover" status="ok" detail=""
 # unix_micros,characteristic_uuid,payload_hex
1784854547341388,ac060002-328c-a28f-9846-5a8aa212661b,01

Header mode/rate fields describe what was requested from the device. They are not treated as observed signal facts. inspect-raw derives packet cadence, payload sizes, command indications, and compatible sample rates from the records themselves. The parser remains backward-compatible with v1 captures.

M2 and M4 Findings

Confirmed M2 notifications contain nine little-endian signed 16-bit pairs:

[channel_1_sample_0, channel_2_sample_0, ... channel_1_sample_8, channel_2_sample_8]

At 33 1/3 packets/s, that is 300 samples/s/channel. Controlled contact-removal tests identify channel 1 as lead I and channel 2 as lead II. Standard limb-lead relationships derive the remaining four leads.

M4 was probed as a nominal dual-lead 600 Hz request. On the tested device it returned indication 0x03, but still produced 36-byte packets at 33.336 packets/s that decode cleanly as the same 300 Hz M2-compatible transport. The software therefore records M4 raw data but does not label or export it as 600 Hz.

Workspace

Path Responsibility
crates/kardia-core Protocol-neutral samples, timestamps, recording metadata, and limb-lead derivation
crates/kardia-ble Device identification, mode commands, live BLE capture, raw capture parsing, and packet decoding
crates/kardia-cli Scan/probe/capture/inspect/export workflows and live terminal view
docs/re Sanitized protocol evidence, hypotheses, and unresolved questions
scripts/bleak_probe.py Independent Python/CoreBluetooth probe for comparing bleak with btleplug

The BLE layer owns device evidence and lossless capture. The core crate remains independent of AliveCor-specific framing so future UI, replay, and model code can consume stable ECG types.

Cross-checking with Bleak

The Python probe runs the same scan/read/subscribe/unlock sequence through Apple CoreBluetooth via bleak:

uv venv .venv
uv pip install --python .venv/bin/python bleak
.venv/bin/python scripts/bleak_probe.py --stream --scan-secs 120

If a step works in bleak but not btleplug, investigate the Rust BLE layer. If both fail at the same operation, investigate bonding or device behavior.

Development

cargo fmt --all -- --check
cargo check --workspace
cargo test --workspace
cargo clippy --workspace --all-targets -- -D warnings

The workspace currently has three crates and requires no external services for unit tests. Live BLE verification requires the physical device.

Contributing

Reverse-engineering changes should:

  • preserve raw bytes and receive timestamps before interpretation;
  • distinguish requested configuration from observed behavior;
  • cite the device, firmware, host, command, and capture statistics;
  • add parser/decoder tests with synthetic or deliberately anonymized fixtures;
  • avoid committing personal ECG captures; and
  • update docs/re/kardia-6l.md when evidence changes an assumption.

Please run the complete development command set before opening a pull request.

Privacy

ECG captures are health-related data and may also contain device identifiers. The repository ignores captures/ by default. Do not publish a capture unless it has been intentionally reviewed, minimized, and anonymized.

When sharing a generated report, replace its local input path with a neutral label:

cargo run -p kardia-cli -- render-ecg \
  captures/kardia-m2.csv \
  --source-label "Anonymized research capture" \
  --out report.svg

License

Licensed under either the Apache License, Version 2.0 or the MIT License, at your option, as declared by the workspace manifests.