An independent, open audit of the robustness and fairness of physiological-signal AI.
FairPulse evaluates how machine-learning models built on photoplethysmography (PPG) — the optical signal behind smartwatches and pulse oximeters — hold up when the device, the conditions, or the patient population (notably skin tone) change. It is not another PPG model: it is a neutral arbiter that takes existing open models and stress-tests them on one honest, leakage-aware protocol. A benchmark suite, not a single-score leaderboard.
Status — v1. Robustness audit on open data and a skin-tone fairness audit executed on OpenOximetry (PhysioNet restricted, used locally under a signed DUA, never redistributed). The headline finding is honest and two-level — see Results.
git clone https://github.com/sahimhiba-debug/fairpulse && cd fairpulse
uv sync # core deps only (no torch needed for the demo)
uv run python scripts/00_demo_audit.py # full pipeline on SYNTHETIC data -> a real reportThis runs the complete audit (frozen-embedding probe → per-skin-group MAE → subject-bootstrap CIs
→ leave-one-skin-tone-out) on bundled synthetic fixtures and writes docs/demo_audit_report.md.
No restricted data, no model weights, nothing under a DUA — it proves the pipeline and shows the report
shape. (The numbers are illustrative, not real model performance.)
The zoo interface is one method: embed(windows, fs) -> (N, embedding_dim). Wrap your model in it,
then reuse the same pipeline:
import numpy as np
from fairpulse.demo import run_demo_audit # synthetic data, swap the model
from fairpulse.benchmark.evaluate import audit_embeddings, embed_segments
class MyModel:
name = "my-ppg-model"
embedding_dim = 256
def embed(self, windows: np.ndarray, fs: int) -> np.ndarray: # windows: (N, L) raw PPG @ fs Hz
x = my_preprocess(windows, fs) # your resample / normalize
return my_encoder(x) # -> (N, 256) numpy
# (a) quickest — drop your model into the synthetic demo:
report = run_demo_audit(model=MyModel())
# (b) on YOUR data — build a list[fairpulse.data.base.Segment] (signal, fs, subject_id,
# skin_group, label), then:
# embs = embed_segments(MyModel(), segments)
# report = audit_embeddings(embs, labels, subjects, skin_groups=groups) # MAE/group + CIs + fairnessSplits are always at the patient / device / skin-tone level (never the segment level), with
subject-level bootstrap confidence intervals. See src/fairpulse/zoo/ for real adapters
(PaPaGei-S, Pulse-PPG) as worked examples.
- Model zoo — uniform 512-d adapters for PaPaGei-S and Pulse-PPG (encoders vendored and
attributed, see
THIRD_PARTY_NOTICES.md); Tier-1 wrappers prepared for AnyPPG, SIGMA-PPG, PPG-VascularAge (weights not bundled). - Leakage-aware protocol — frozen-embedding Ridge probe; LOSO / leave-one-device-out / leave-one-skin-tone-out; subject-bootstrap CIs.
- Robustness audit (open data) — per-condition breakdown + cross-dataset transfer.
- Fairness audit (OpenOximetry, local) — per-skin-group error + the data-level SpO2↔SaO2 bias and occult-hypoxemia rate.
- Audit tool (stretch) — a FastAPI endpoint that takes embeddings and returns a report.
Figure 2 — Leakage-aware audit protocol: frozen-embedding Ridge probe, patient/device/skin-tone splits (never segment-level), subject-bootstrap CIs, LOSTO.
Robustness (open data, heart-rate MAE in bpm, LOSO ± 95% CI): Pulse-PPG leads on these wrist/finger
sets (field-trained), and physical effort ~doubles the error for both models — a robustness
observation, not a winner. Full tables in docs/rapport_v1.md.
Fairness by skin tone (OpenOximetry) — read at TWO levels:
- Data level (model-independent) — robust & significant. The pulse oximeter overestimates true arterial saturation more for darker skin: SpO2−SaO2 bias light 0.67 → dark 2.21 pts (CIs disjoint), occult hypoxemia ~3% (dark) vs ~0% (light). The documented bias (Sjoding 2020, Fawzy 2022) is reproduced.
- Model level — same direction, NOT conclusive. Both PaPaGei-S and Pulse-PPG show higher saturation-estimation error on dark skin (SpO2 MAE dark ~2.4 vs light ~1.9), but the dark group is small (11 patients with usable PPG) → wide, overlapping CIs.
Numbers, CIs and the honest reading are in docs/rapport_v1.md §5.4;
machine-readable in docs/equity_results_openox.json.
Figure 1 — Data-level pulse-oximeter bias (SpO₂−SaO₂) by skin tone, OpenOximetry. Model-independent; 95% CI (subject-bootstrap). dark n=13 (small sample).
- Open, downloadable (robustness): PPG-DaLiA (CC-BY-4.0), BIDMC (ODC-BY) —
scripts/02_fetch_assets.py. - OpenOximetry (fairness): PhysioNet restricted — requires a signed DUA. It stays local
in
data/raw/(gitignored), is never committed/pushed, and never sent to any online service (including the FastAPI endpoint). You download it yourself; seescripts/openoximetry_download.md. - Synthetic fixtures (no DUA): bundled in
src/fairpulse/demo.py— the Quickstart path.
No raw data, model weights, or secrets are ever committed (.gitignore enforced).
Honest by design — full table in docs/rapport_v1.md §6:
- Skin-tone fairness: done on OpenOximetry, but the dark group is small (13 metadata / 11 with PPG) → the model-level gap is suggestive, not conclusive; widening it means combining datasets.
- Leave-one-device-out for a PPG model: not possible even with OpenOximetry — its raw PPG comes from a single reference recorder; the 53 oximeters vary the label, not the model input.
- Saturation labels are encounter-level (the v1.1.1 PPG is not time-synced to SpO2/SaO2 samples).
- A suite, not a leaderboard. Existing models are subjects of audit, not baselines to beat.
- Splits at patient / device / skin-tone level; report leave-one-out with confidence intervals.
- Code is MIT (see
LICENSE). Datasets and model weights keep their own licenses and access terms. PaPaGei weights carry no declared license →to_verify, local research use only, never redistributed (seeTHIRD_PARTY_NOTICES.md). - v1 is research / evaluation tooling, not a medical device.
src/fairpulse/
zoo/ uniform inference adapters (PaPaGei-S, Pulse-PPG -> 512-d) + _vendor encoders
data/ loaders: PPG-DaLiA, BIDMC, OpenOximetry; common Segment schema
benchmark/ leakage-aware splits, Ridge probe, metrics, subject-bootstrap CIs, fairness
audit/ report rendering + FastAPI endpoint (stretch)
demo.py synthetic fixtures + DemoModel (the Quickstart path)
scripts/ 00_demo_audit (fixtures) · 02_fetch · 03_run_audit (open) · 04 equity · 05 integrity
docs/ rapport_v1.md (results + gaps map), catalogs, check logs
59 tests, ruff clean. Report figures are checked against the audit JSON. Raw data, weights and secrets
are never committed.
Citation. If you use FairPulse in your research, please cite:
Sahim Elmir, H. (2026). FairPulse: an independent robustness & fairness audit suite for physiological-signal AI (PPG) (v1.0.0). Zenodo. https://doi.org/10.5281/zenodo.20723318
A CITATION.cff file is also included for citation managers.
FairPulse audits, and is grateful to, the open work it builds on — PaPaGei (Nokia Bell Labs),
Pulse-PPG, the OpenOximetry repository, and the public datasets PPG-DaLiA and BIDMC.
See THIRD_PARTY_NOTICES.md for attribution and license status.